For Laboratory & Research Use Only — Not for Human or Veterinary Use
United Peptides
Quality and testing

ESI vs MALDI: What Mass Spectrometry Confirms, and What It Cannot

September 26, 20265 min readUnited Peptides

  • mass spectrometry
  • identity
  • analytical chemistry
A fine blue spray from a glass needle toward a metal cone

Mass spectrometry weighs a molecule; it does not read it. That one sentence covers most of what an identity panel can and cannot prove. The two instruments a peptide certificate will name, electrospray and MALDI, both produce a mass, and they get there by different routes that leave different fingerprints on the result. Knowing which was used tells you what the number means and which errors it could hide.

Two ways to get a peptide into the gas phase

A mass spectrometer measures the mass-to-charge ratio of ions in a vacuum. The difficulty with peptides is getting a large, fragile, polar molecule out of solution and into that vacuum as an intact ion. Two techniques solved it, and they share a Nobel prize.

Electrospray (ESI)MALDI
Sample stateSolution, sprayed through a charged needleDried into a crystalline matrix, hit with a laser
Charge on the ionsMultiple: +2, +3, +4…Mostly single: +1
Couples to HPLCYes, directly (LC-MS)No; run separately
What the spectrum showsA ladder of peaks, one per charge stateOne main peak per component
Best atImpurity identification alongside separationFast identity check of a purified sample

Why does an electrospray spectrum show several peaks for one peptide?

Because the peptide picks up several protons on the way in, and each charge state appears at a different position. A 4,000 Da peptide carrying three protons appears at roughly 1,334 on the m/z axis; with four protons, at about 1,001. The spectrum is a ladder, and software deconvolutes it back to a single neutral mass. That is the number on the certificate. A raw ESI spectrum therefore looks nothing like a single peak, and a reader expecting one may think something is wrong when the instrument is behaving exactly as designed.

Why is MALDI mostly singly charged?

Because of how the ions form. The sample is embedded in a matrix that absorbs the laser, and the energy that vaporises it typically transfers one proton to each peptide. The spectrum shows the molecule at its mass plus one, with a smaller peak at mass plus twenty-three where sodium has attached instead. It is simpler to read than electrospray and faster to run, which makes it the common choice for confirming that a purified batch is the intended molecule. What it cannot easily do is separate components first, which is why it is a poorer tool for finding out what an impurity is.

Monoisotopic and average mass

This is the detail most likely to cause confusion when comparing a certificate to a calculation. Every peptide contains carbon, and about one carbon in ninety is the heavier carbon-13 isotope. A molecule with 200 carbons will therefore appear as a cluster of peaks one mass unit apart, depending on how many heavy atoms it happened to contain.

The monoisotopic mass is the peak with no heavy isotopes — the lightest in the cluster. The average mass weights all of them by abundance. For a 4,000 Da peptide the two differ by about 2 Da, and a certificate that reports one against a calculation made in the other will look like an error of two units when nothing is wrong. High-resolution instruments resolve the cluster and report monoisotopic; lower-resolution ones see a single broad peak at the average. The certificate should say which, and the method line is where it says it.

How close should the observed mass be to the calculated one?

On a modern instrument, within a fraction of a dalton once both numbers are on the same basis. A discrepancy of exactly one unit is usually a deamidation or an amidation state; of two, monoisotopic against average; of sixteen, an oxidised methionine; of eighteen, a ring closure or a water loss; of forty-two, an acetyl group. Each of these is a specific, checkable explanation. A discrepancy that matches none of them is a reason to ask what was analysed. The modifications written around a sequence account for most of them in advance.

What mass cannot tell you

Three things, and each is a real limitation on the identity panel.

Sequence order. The same residues in a different order weigh the same. Mass confirms composition; only fragmentation — tandem MS, where the ion is broken and the pieces weighed — reads order, and most routine certificates do not include it.

Stereochemistry. An L residue and its D mirror image weigh the same. A diastereomer impurity is invisible by mass and needs chiral analysis.

Connectivity. A peptide with two disulphide bridges has three possible ways to pair its four cysteines, all at the same mass. Which one formed needs a mapping experiment, not a weighing.

What does LC-MS add over either alone?

A mass for every peak on the chromatogram, in one run. The column separates the components and the electrospray source weighs each as it elutes, so an anonymous 1.2% impurity on a purity trace becomes a peak with a mass, and a mass sixteen units below the main one is an identified deletion sequence rather than a number. It is the most informative single analysis a routine certificate can carry, and a certificate that reports impurities by identity rather than by percentage has almost certainly used it.

Is a mass match enough to trust a batch?

It is enough to trust that the batch is the intended composition, which is a great deal and is not everything. It does not establish purity, which is the chromatogram's job; it does not establish sequence order, handedness or bridge connectivity, which need dedicated experiments; and it does not establish how much of the vial is peptide, which is net content. Each panel answers one question. Mass answers "is this the right molecule" to a high standard and answers nothing else, and reading it as more than that is where certificates get over-interpreted.

All products referenced here are supplied for laboratory and research use only. They are not drugs, foods, supplements or cosmetics, and are not for human or veterinary use.

Keep reading

More from the research blog